- PII
- S30344980S0130308225100071-1
- DOI
- 10.7868/S3034498025100071
- Publication type
- Article
- Status
- Published
- Authors
- Volume/ Edition
- Volume / Issue number 10
- Pages
- 68-74
- Abstract
- The paper describes a practically implemented method for correcting the geometrical aberrations introduced by a scanning device as a proposed stage of the computer radiography system qualification. The presence of geometrical distortions in the scanned images affects the metrological characteristics of the measurement methods and techniques that are using these systems. The method allows for correction of a systematic error obtained when scanning photostimulable phosphor detectors on the digital radiography devices. The main stages of method implementation include fabrication of the calibration sample, conducting reference instrumental measurements, comparing reference measurements with digital image processing results, and estimation and correcting the errors. During the period 2022 to 2024, the examination results for three scanning devices were analyzed. The use of the geometrical aberration correction method allows for minimization of distortions as well as estimation of quality and stability of the scanning devices.
- Keywords
- геометрические аберрации система цифровой радиографии сканирующие устройства рентгеновское изображений фотолюминесцентные экраны с запоминанием градуировка рентгенографические измерения
- Date of publication
- 01.09.2025
- Year of publication
- 2025
- Number of purchasers
- 0
- Views
- 13
References
- 1. Майоров А.А. Компьютерная радиография с использованием флуоресцентных запоминающих пластин — что это такое? // В мире неразрушающего контроля. 2004. № 3. С. 42—43.
- 2. Троицкий В.А. Новые возможности радиационного контроля качества сварных соединений // Автоматическая сварка. 2015. № 7. С. 56—60.
- 3. Троицкий В.А., Михайлов С.Р., Пастовенский Р.О., Шило Д.С. Современные системы радиационного неразрушающего контроля // Техническая диагностика и неразрушающий контроль. 2015. № 1. С. 23—35.
- 4. Logan Clint M., Martz Harry E., Schneberk Daniel J., Shull Peter J. X-Ray Imaging. Fundamentals, Industrial Techniques and Applications. 2017.
- 5. Seibert (Chair) J. Anthony, Bogucki Terese M., Ciona Ted, Huda Walter, Karellas Andrew, Mercier John R., Samet Elisan, Shepard S. Jeff, Stewart Brent K., Strauss Keith J., Suleiman Orhan H., Tucker Doug, Uzenoff Robert A., Weiser John Conrad. Acceptance Testing and Quality Control of Photosimulable Storage Phosphor Imaging Systems. Report No. 93 of American Association of Physicists in Medicine Task Group 10. 2006. 68 с.
- 6. Недавний О.И., Удод В.А. С Современное состояние систем цифровой рентгенографии (обзор) // Дефектоскопия. 2001. № 8. С. 62—82.
- 7. Степанов А.В., Косарина Е.И., Демидов А.А. Компьютерная рентгенография с применением фотостатулированных пластин // Авиационные материалы и технологии. 2015. № 4. С. 79—85.
- 8. Curcio A., Andreoli P., Cipriani M., Claps G., Consoli F. Imaging plates calibration to X-rays — Plasma Physics by Laser and Applications / ENEA Research Centre, Frascati, Italy. 2016. 14 с.
- 9. Rosenberg M.J., Thorn D.B., Izumi N., Williams D., Rowland M., Torres G., Haugh M., Hillyard P., Adelman N., Schuler T., Barrios M.A., Holder J.P., Schneider M.B., Fournier K.B., Bradley D.K., Regan S.P. Image-plate sensitivity to x rays at 2 to 60 keV / Review of Scientific Instruments 90. 013506. 2019.
- 10. Потрахов Н.Н. Метод и особенности формирования теневого рентгеновского изображения микрофокусными источниками излучения // Вестник новых медицинских технологий. 2007. Т. XIV. № 3. С. 167—169.
- 11. Kolesnikov P.A., Politov V.Yu., Li E.S., Yaskevich A.P., Kolesnikov S.A., Ekimov Yu.M., Malyshev M.A., Akhmetov A.R., Petrov D.V., Nikitin O.A., Perzin I.V., Trunev Yu.A., Alukhanov M.G., Burdakov A.V., Danilov V.V., Kurkachenkov V.V., Popov S.S., Skovorodin D.I., Zhivankov K.I. Technique for measuring of LIA spot size on target by X-rays propagating in beam backward direction // Journal of Instrumentation. 2020. JINST 15. P. 10018.
- 12. Schach von Wittman A.E., Logan C.M., Rikard R.D. Using a tungsten rollbar to characterize the source spot of a megavoltage bremsstrahlung linac // The International Journal of Medical Physics Research and Practice, Lawrence Livermore National Laboratory, Livermore, California 94550. 2002. No. 8. P. 1797—1806.
- 13. Екимов Ю.М., Малышев М.А., Никитин О.А., Сергеев А.В. Способ градуировки распределения пространственного разрешения устройств получения цифрового рентгеновского изображения / Патент № 2826349 от 09.09.2024.
- 14. Августинович К.А. Основы фотографической метрологии. М.: Легирокосбытиздат, 1990.
- 15. Екимов Ю.М., Малышев М.А., Никитин О.А. Свидетельство о государственной регистрации программы для ЭВМ «CRCalibrationUtility» № 2022681907 от 17.11.2022.
- 16. Шешин Е.П., Еременко А.Г., Куров И.О., Кузнецова А.А., Нечепуренко Е.В., Сапунов М.А. Калибровка системы стереоскопического компьютерного зрения // Труды МФТИ. 2018. Т. 10. № 2. С. 80—89.
- 17. Демин Н.С., Ильясова Н.Ю., Устинов А.В. Разработка алгоритма коррекции искажений центрального проецирования на рентгеновских изображениях / Информационные технологии и нано-технологии - 2017. Компьютерная оптика и нанофотоника. 2017. С. 34—39.
- 18. Герасимов С.И., Костин В.И. Калибровка неметрических цифровых фото- и видеокамер для фотографических измерений / Пособие для студентов вузов. Саров, 2015.